Date: Wed, 29 May 2024 07:42:10 -0500 (CDT)Message-ID: <1923523092.901.1716986530641@manuals.iodp.tamu.edu>Subject: Exported From ConfluenceMIME-Version: 1.0Content-Type: multipart/related; boundary="----=_Part_900_1045014642.1716986530641"------=_Part_900_1045014642.1716986530641Content-Type: text/html; charset=UTF-8Content-Transfer-Encoding: quoted-printableContent-Location: file:///C:/exported.html

Author:

E. Fisher

Reviewer(s):

Editor(s):

S. Boehm, S. Frank

Management Approval (Name, Ti=tle, Date):

B. Mills (Lab Officer)

Audience:

Technicians

Origination date:

<=/td>

4/30/2013

Current version:

=td>

7/30/2018

Revised:

Domain:

Thin Section Lab

System:

Logitech LP50

Keywords:

Lapping Samples

This user guide will help thin section technicians bring their sample to= 40 microns thick. This will prepare the sample for the final stage of poli=shing on the Logitech WG2. For more information refer to the Logitech Thin Rock Section Production =manual, available in the Thin Section Lab both on a Logitech thumb drive in= the bookcase and computer under the "Manual" folder on the desktop. Or, re=fer to Operation and Maintenanc=e of the LP50 Lapping and Polishing Machine manual, located in the b=lack binder in the Thin Section Lab cabinet labeled "LP50, PM5 and Associat=ed Equipment".

SETTING THE JIG

  1. Use the jig labeled "Lap".
  2. Clean jig and Logitech scale by wiping them with a kimwipe and isoprop=yl alcohol.=20
    1. Refer to the Logitech LP50 Set Up user guide for more information on= scale use.
  3. Gently set the scale on the jig (Fig. 1) and wiggle the scale.
  4. The jig should be set to 1201.=20
    1. Use the middle gauge on the scale.=20
      1. The small dial (Fig. 1) on the bottom left side should read 12.
      2. The large face (Fig. 1) should read 01.=20
        1. If it is not set to 1201, change it by inserting the Tommy bar i=nto the holes on the main post in the middle of the jig (Fig. 1).=20
          1. Rotating the Tommy bar clockwise will lower the chuck face cau=sing the numbers on the scale to increase.
          2. Rotating the Tommy bar counterclockwise will raise the chuck f=ace, causing the numbers on the scale to decrease.


LAPPING SAMPLES ON LP50 (1)
&nb=sp; &=nbsp; = &=nbsp;&nbs=p;&=nbsp; Figure 1

5. Remove the scale after the jig has been set.

LAPPING SAMPLES

  1. Clean samples and the chuck face with kimwipe and isopropyl alcohol.=li>
  2. Set six clean samples (unfrosted side down) on the chuck face over the= vacuum grooves. There should be no spaces between any of the slides.=20
    1. If there are 1 or 2 samples, place them in the middle of the chuck f=ace and frosted slides on the outer sides.=20
      1. The same frosted slides can used be as "filler slides" throughout =the expedition.
    2. If there are 3 or 4 samples, place them on the outer sides of the ch=uck face and the filler slides in the middle (Fig. 2).


LAPPING SAMPLES ON LP50 (2)
&nbsp=; &nb=sp; &nbsp=;&n=bsp; Figure 2=em>

3. Connect vacuum hose, if not already connected.

4. Check oil level in the vacuum pump under the counter.

5. Turn on vacuum pump with the button labeled Vacuum u=nder the joystick on the LP50.

6. Close valve on the LP50 to the left of the vacuum gauge (Fig. 3).

7. The vacuum gauge should read less than 150 mbar (Fig. 3). It should b=e in the green portion of the gauge.


LAPPING SAMPLES ON LP50 (3)

&=nbsp; &nb=sp;=&nb=sp; &nbsp=;Figure 3

    1. If there is not a proper vacuum seal right away, try pressing lightl=y on the slides to help them seal and make sure the exhaust flap (Fig. 3) i=s clean.
    2. If that does not work, take off the slides and clean them and chuck =face again.

8. Gently set the jig upright on the lapping plate in the forward arm.=p>

9. The plate monitor can stay in place on the back arm.

10. Turn on the abrasive drum by overriding and make sure it is dripping= properly.

11. Reset timer and set it to 10 to 20 minutes depending how hard the sa=mples are.

    1. Samples that cut quickly on the PetroThin, lap for about 10 to 12 mi=nutes for the first time.
    2. Samples that cut slowly on the PetroThin, lap for about 15 to 20 min=utes for the first time.

12. Set the plate speed to 5 rpms.

13. When the plate is wet, press Start.

14. Increase the plate speed in small increments to 58 rpms.

15. The jig should start to spin after a couple minutes when all the sam=ples begin to even out to the same thickness.

    1. Bungee cords can be added to the vacuum hose to assist the jig spin.=
    2. Do not walk out of the lab when the LP 50 is running.
    3. If the samples are left on too long, the abrasive will "pluck" miner=als out of the sample, leaving a hole.

16. When time is up, press Stop.

    1. It is best to bring the speed back down to 5 rpms after the lapping =plate is stopped to help prevent starting it back up at a high speed.

17. Take the jig off the plate carefully.

18. Open vacuum valve and turn it off.

19. Place jig chuck face up in the sink.

20. Scrub the jig and slides gently with a soft brush to break the vacuu=m seal.

21. Clean the jig by scrubbing it down with warm water and a soft brush.= Then, blow air through the jig and on the chuck face with the air gun.

22. Rinse the samples with warm water, and then wipe them with a kimwipe= and isopropyl alcohol. Blow them dry with the air gun.

MEASURING SAMPLES

  1. Hold the micrometer with your pinkie and ring finger through the "C" p=art of the micrometer. Your index finger and thumb will move the dial.
  2. Clean the micrometer before use by placing a piece of paper in between= the two bars and closing it gently (do not apply force). Pull the paper ou=t and repeat.=20
    1. Gently close the micrometer all the way. It should read zero.
  3. To measure the sample, place it in between the two bars and close them= gently onto the slide where the sample sits.=20
    1. Take a measurement at the middle and sides to make sure that the sam=ple is even all the way across.
  4. On the micrometer, each tick mark is 10 um. The difference between the= thickness you measure and the thickness of the slide is the thickness of t=he sample. For an example, if you measure 19 on the micrometer and the slid=e is 15, that mean the sample is 40 um.=20
    1. Soft samples should be about 45 to 50 um.=20
      1. If the sample is too thin, they will either not get a good polish =before they reach their target thickness of 30 um, or, they will become too= thin.
    2. Medium or hard samples should be about 40 um. This is the majority o=f samples.
    3. Very hard samples should be about 35 um.=20
      1. If the sample is too thick, it will take a very long time to polis=h down to the target thickness.
  5. Write the thickness next to each sample on the sample tray.
  6. If the samples are too thick, put them back on the lapping wheel for a= couple more minutes.

Original version created by Emily Fisher

4/30/2013

  • X397T archived version
  • LAPPING SAMPLES ON THE LOGITECH LP50 374.docx
LAPPING SAMPLES ON LP50 (2024)

References

Top Articles
Latest Posts
Article information

Author: Tish Haag

Last Updated:

Views: 5887

Rating: 4.7 / 5 (67 voted)

Reviews: 90% of readers found this page helpful

Author information

Name: Tish Haag

Birthday: 1999-11-18

Address: 30256 Tara Expressway, Kutchburgh, VT 92892-0078

Phone: +4215847628708

Job: Internal Consulting Engineer

Hobby: Roller skating, Roller skating, Kayaking, Flying, Graffiti, Ghost hunting, scrapbook

Introduction: My name is Tish Haag, I am a excited, delightful, curious, beautiful, agreeable, enchanting, fancy person who loves writing and wants to share my knowledge and understanding with you.